258 lines
16 KiB
JSON
258 lines
16 KiB
JSON
{
|
||
"dataType": "CVE_RECORD",
|
||
"dataVersion": "5.2",
|
||
"cveMetadata": {
|
||
"cveId": "CVE-2025-14097",
|
||
"assignerOrgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
|
||
"state": "PUBLISHED",
|
||
"assignerShortName": "Radiometer",
|
||
"dateReserved": "2025-12-05T10:50:03.683Z",
|
||
"datePublished": "2025-12-17T12:36:03.779Z",
|
||
"dateUpdated": "2025-12-17T14:41:13.252Z"
|
||
},
|
||
"containers": {
|
||
"cna": {
|
||
"affected": [
|
||
{
|
||
"defaultStatus": "unaffected",
|
||
"product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
|
||
"vendor": "Radiometer Medical Aps",
|
||
"versions": [
|
||
{
|
||
"status": "affected",
|
||
"version": "Application software versions < 3.5MR11 with Windows 7, Windows XP operating systems",
|
||
"versionType": "ABL90 DMS (Data Management system)"
|
||
},
|
||
{
|
||
"status": "unaffected",
|
||
"version": "Application software versions >= 3.5MR11 with Windows 10 operating system",
|
||
"versionType": "ABL90 DMS (Data Management system)"
|
||
}
|
||
]
|
||
},
|
||
{
|
||
"defaultStatus": "unaffected",
|
||
"product": "ABL800 BASIC and ABL800 FLEX Analyzers",
|
||
"vendor": "Radiometer Medical Aps",
|
||
"versions": [
|
||
{
|
||
"status": "affected",
|
||
"version": "Application software versions < 6.20MR2 with Windows 7, Windows XP operating systems",
|
||
"versionType": "ABL800 DMS(Data Management System)"
|
||
},
|
||
{
|
||
"status": "unaffected",
|
||
"version": "Application software versions >= 6.20MR2 with Windows 10 operating system",
|
||
"versionType": "ABL800 DMS(Data Management System)"
|
||
}
|
||
]
|
||
},
|
||
{
|
||
"defaultStatus": "unaffected",
|
||
"product": "AQT90 FLEX Analyzers",
|
||
"vendor": "Radiometer Medical Aps",
|
||
"versions": [
|
||
{
|
||
"status": "affected",
|
||
"version": "Application software versions <= 8.13 MR2 with Windows 7, Windows XP operating systems",
|
||
"versionType": "AQT90 DMS (Data Management System)"
|
||
},
|
||
{
|
||
"status": "unaffected",
|
||
"version": "Application software versions >= 8.13 MR2 with Windows 10 operating system",
|
||
"versionType": "AQT90 DMS (Data Management System)"
|
||
}
|
||
]
|
||
}
|
||
],
|
||
"configurations": [
|
||
{
|
||
"lang": "en",
|
||
"supportingMedia": [
|
||
{
|
||
"base64": false,
|
||
"type": "text/html",
|
||
"value": "Affected application software version is in use AND Remote support feature is enabled in the Analyzer.\n\n<br>"
|
||
}
|
||
],
|
||
"value": "Affected application software version is in use AND Remote support feature is enabled in the Analyzer."
|
||
}
|
||
],
|
||
"credits": [
|
||
{
|
||
"lang": "en",
|
||
"type": "reporter",
|
||
"value": "Florian Hauser and Fabian Weber from CODE WHITE GmbH"
|
||
}
|
||
],
|
||
"datePublic": "2025-12-17T12:35:00.000Z",
|
||
"descriptions": [
|
||
{
|
||
"lang": "en",
|
||
"supportingMedia": [
|
||
{
|
||
"base64": false,
|
||
"type": "text/html",
|
||
"value": "A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. <span style=\"background-color: rgb(255, 255, 255);\">Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.</span>\n\n<br><br>Required Configuration for Exposure: <span style=\"background-color: rgb(255, 255, 255);\">Affected application software version is in use and remote support feature is enabled in the analyzer. </span>Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution:<br>Customers should ensure the following:<br>• The network is secure, and access follows best practices.<br>Local Radiometer representatives will contact all affected customers to discuss a permanent solution. <br>Exploit Status:<br>\n\n<span style=\"background-color: rgb(255, 255, 255);\">Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.</span>\n\n<br>"
|
||
}
|
||
],
|
||
"value": "A vulnerability in the application software of multiple Radiometer products may allow remote code execution and unauthorized device management when specific internal conditions are met. Exploitation requires that a remote connection is established with additional information obtained through other means. The issue is caused by a weakness in the analyzer’s application software. Other related CVE's are CVE-2025-14095 & CVE-2025-14096. Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\nRequired Configuration for Exposure: Affected application software version is in use and remote support feature is enabled in the analyzer. Temporary work Around: If the network is not considered secure, please remove the analyzer from the network. Permanent solution:\nCustomers should ensure the following:\n• The network is secure, and access follows best practices.\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution. \nExploit Status:\n\n\nResearchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication."
|
||
}
|
||
],
|
||
"exploits": [
|
||
{
|
||
"lang": "en",
|
||
"supportingMedia": [
|
||
{
|
||
"base64": false,
|
||
"type": "text/html",
|
||
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication.\n\n<br>"
|
||
}
|
||
],
|
||
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any publicly available exploits at the time of this publication."
|
||
}
|
||
],
|
||
"impacts": [
|
||
{
|
||
"capecId": "CAPEC-555",
|
||
"descriptions": [
|
||
{
|
||
"lang": "en",
|
||
"value": "CAPEC-555: Remote Services with Stolen Credentials"
|
||
}
|
||
]
|
||
},
|
||
{
|
||
"capecId": "CAPEC-653",
|
||
"descriptions": [
|
||
{
|
||
"lang": "en",
|
||
"value": "CAPEC-653: Use of Known Operating System Credentials"
|
||
}
|
||
]
|
||
}
|
||
],
|
||
"metrics": [
|
||
{
|
||
"cvssV3_1": {
|
||
"attackComplexity": "LOW",
|
||
"attackVector": "NETWORK",
|
||
"availabilityImpact": "HIGH",
|
||
"baseScore": 7.2,
|
||
"baseSeverity": "HIGH",
|
||
"confidentialityImpact": "HIGH",
|
||
"integrityImpact": "HIGH",
|
||
"privilegesRequired": "HIGH",
|
||
"scope": "UNCHANGED",
|
||
"userInteraction": "NONE",
|
||
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:H/UI:N/S:U/C:H/I:H/A:H",
|
||
"version": "3.1"
|
||
},
|
||
"format": "CVSS",
|
||
"scenarios": [
|
||
{
|
||
"lang": "en",
|
||
"value": "GENERAL"
|
||
}
|
||
]
|
||
}
|
||
],
|
||
"problemTypes": [
|
||
{
|
||
"descriptions": [
|
||
{
|
||
"cweId": "CWE-287",
|
||
"description": "CWE-287: Improper Authentication",
|
||
"lang": "en",
|
||
"type": "CWE"
|
||
}
|
||
]
|
||
}
|
||
],
|
||
"providerMetadata": {
|
||
"orgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
|
||
"shortName": "Radiometer",
|
||
"dateUpdated": "2025-12-17T13:14:55.334Z"
|
||
},
|
||
"references": [
|
||
{
|
||
"url": "https://www.radiometer.com/myradiometer"
|
||
}
|
||
],
|
||
"solutions": [
|
||
{
|
||
"lang": "en",
|
||
"supportingMedia": [
|
||
{
|
||
"base64": false,
|
||
"type": "text/html",
|
||
"value": "Customers should ensure the following:<br>• The network is secure, and access follows best practices.<br>Local Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\n\n\n<br>"
|
||
}
|
||
],
|
||
"value": "Customers should ensure the following:\n• The network is secure, and access follows best practices.\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution."
|
||
}
|
||
],
|
||
"source": {
|
||
"discovery": "EXTERNAL"
|
||
},
|
||
"timeline": [
|
||
{
|
||
"lang": "en",
|
||
"time": "2025-12-17T12:35:00.000Z",
|
||
"value": "CVE Published"
|
||
}
|
||
],
|
||
"title": "Remote Code Execution Vulnerability in Radiometer Products",
|
||
"workarounds": [
|
||
{
|
||
"lang": "en",
|
||
"supportingMedia": [
|
||
{
|
||
"base64": false,
|
||
"type": "text/html",
|
||
"value": "If the network is not considered secure, please remove the analyzer from the network."
|
||
}
|
||
],
|
||
"value": "If the network is not considered secure, please remove the analyzer from the network."
|
||
}
|
||
],
|
||
"x_generator": {
|
||
"engine": "Vulnogram 0.5.0"
|
||
}
|
||
},
|
||
"adp": [
|
||
{
|
||
"metrics": [
|
||
{
|
||
"other": {
|
||
"type": "ssvc",
|
||
"content": {
|
||
"timestamp": "2025-12-17T14:40:30.048229Z",
|
||
"id": "CVE-2025-14097",
|
||
"options": [
|
||
{
|
||
"Exploitation": "none"
|
||
},
|
||
{
|
||
"Automatable": "no"
|
||
},
|
||
{
|
||
"Technical Impact": "total"
|
||
}
|
||
],
|
||
"role": "CISA Coordinator",
|
||
"version": "2.0.3"
|
||
}
|
||
}
|
||
}
|
||
],
|
||
"title": "CISA ADP Vulnrichment",
|
||
"providerMetadata": {
|
||
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
|
||
"shortName": "CISA-ADP",
|
||
"dateUpdated": "2025-12-17T14:41:13.252Z"
|
||
}
|
||
}
|
||
]
|
||
}
|
||
} |