"dataType": "CVE_RECORD",
"dataVersion": "5.2",
"cveMetadata": {
"cveId": "CVE-2025-14096",
"assignerOrgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"state": "PUBLISHED",
"assignerShortName": "Radiometer",
"dateReserved": "2025-12-05T10:50:00.566Z",
"datePublished": "2025-12-17T12:19:27.489Z",
"dateUpdated": "2025-12-17T14:43:16.523Z"
},
"containers": {
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"platforms": [
"Windows XP",
"Windows 7"
],
"product": "ABL90 FLEX and ABL90 FLEX PLUS Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "AQT90 FLEX Analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
},
{
"defaultStatus": "unaffected",
"platforms": [
"Windows 7",
"Windows XP"
],
"product": "ABL800 BASIC and ABL800 FLEX analyzers",
"vendor": "Radiometer Medical Aps",
"versions": [
{
"status": "affected",
"version": "Windows 7 Operating system",
"versionType": "Radiometer Customized Windows Operating System"
},
{
"status": "affected",
"version": "Windows XP Operating system",
"versionType": "Radiometer Customized Windows Operating System"
}
]
}
],
"configurations": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Requires physical access to the Analyzer.\n\n<br>"
}
],
"value": "Requires physical access to the Analyzer."
}
],
"credits": [
{
"lang": "en",
"type": "reporter",
"value": "Florian Hauser and Fabian Weber from CODE WHITE GmbH"
}
],
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.<br><br>Other related CVE's are CVE-2025-14095 &amp; CVE-2025-14097.<br><br>\n\n<span style=\"background-color: rgb(255, 255, 255);\">Affected customers have been informed about this vulnerability. This CVE is being published to provide transparency.</span>\n\n<br><br>\n\n<span style=\"background-color: rgb(255, 255, 255);\">Required Configuration for Exposure:<br></span>\n\n<span style=\"background-color: rgb(255, 255, 255);\">Attacker requires physical access to the analyzer.</span>\n\n<br><br>Temporary work Around:<br>Only authorized people can physically access the analyzer.<br><br>Permanent solution:<br>Local Radiometer representatives will contact all affected customers to discuss a permanent solution.<br><br>Exploit Status:<br>\n\n<span style=\"background-color: rgb(255, 255, 255);\">Researchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.</span>\n\n<br><br>"
}
],
"value": "A vulnerability exists in multiple Radiometer products that allow an attacker with physical access to the analyzer possibility to extract credential information. The vulnerability is due to a weakness in the design and insufficient credential protection in operating system.\n\nOther related CVE's are CVE-2025-14095 & CVE-2025-14097.\n\n\n\nAffected customers have been informed about this vulnerability. This CVE is being published to provide transparency.\n\n\n\n\n\nRequired Configuration for Exposure:\n\n\nAttacker requires physical access to the analyzer.\n\n\n\nTemporary work Around:\nOnly authorized people can physically access the analyzer.\n\nPermanent solution:\nLocal Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\nExploit Status:\n\n\nResearchers have provided a working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"exploits": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication.\n\n<br>"
}
],
"value": "Researchers have provided working proof-of-concept (PoC). Radiometer is not aware of any public exploit code at the time of this publication."
}
],
"impacts": [
{
"capecId": "CAPEC-69",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-69 — Target Programs with Elevated Privileges"
}
]
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "LOCAL",
"availabilityImpact": "HIGH",
"baseScore": 8.4,
"baseSeverity": "HIGH",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-798",
"description": "CWE-798: Use of Hard-coded Credentials",
"lang": "en",
"type": "CWE"
}
]
},
{
"descriptions": [
{
"cweId": "CWE-250",
"description": "CWE-250: Execution with Unnecessary Privileges",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"orgId": "46b595e9-1acc-41cb-9398-adaf98d37a9b",
"shortName": "Radiometer",
"dateUpdated": "2025-12-17T13:03:05.696Z"
},
"references": [
{
"url": "https://www.radiometer.com/myradiometer"
}
],
"solutions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution.\n\n\n\n<br>"
}
],
"value": "Local Radiometer representatives will contact all affected customers to discuss a permanent solution."
}
],
"source": {
"discovery": "EXTERNAL"
},
"timeline": [
{
"lang": "en",
"time": "2025-12-17T13:01:00.000Z",
"value": "CVE Published"
}
],
"title": "Credential Disclosure vulnerability in Radiometer Products",
"workarounds": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "Only authorized people can physically access the analyzer.\n\n\n\n<br>"
}
],
"value": "Only authorized people can physically access the analyzer."
}
],
"x_generator": {
"engine": "Vulnogram 0.5.0"
}
},
"adp": [
{
"metrics": [
{
"other": {
"type": "ssvc",
"content": {
"timestamp": "2025-12-17T14:42:57.843796Z",
"id": "CVE-2025-14096",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"version": "2.0.3"
}
}
}
],
"title": "CISA ADP Vulnrichment",
"providerMetadata": {
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP",
"dateUpdated": "2025-12-17T14:43:16.523Z"
}
}
]
}
}