"dataType": "CVE_RECORD",
"dataVersion": "5.1",
"cveMetadata": {
"cveId": "CVE-2024-9832",
"assignerOrgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
"state": "PUBLISHED",
"assignerShortName": "Baxter",
"dateReserved": "2024-10-10T19:24:46.919Z",
"datePublished": "2024-11-14T21:03:16.721Z",
"dateUpdated": "2024-11-18T15:37:00.311Z"
},
"containers": {
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "Life2000 Ventilation System",
"vendor": "Baxter",
"versions": [
{
"status": "affected",
"version": "06.08.00.00 and prior"
}
]
}
],
"datePublic": "2024-11-14T21:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure.<br>"
}
],
"value": "There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure."
}
],
"impacts": [
{
"capecId": "CAPEC-49",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-49 Password Brute Forcing"
}
]
},
{
"capecId": "CAPEC-441",
"descriptions": [
{
"lang": "en",
"value": "CAPEC-441 Malicious Logic Insertion"
}
]
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "LOW",
"attackVector": "LOCAL",
"availabilityImpact": "HIGH",
"baseScore": 9.3,
"baseSeverity": "CRITICAL",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "CHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-307",
"description": "CWE-307 Improper Restriction of Excessive Authentication Attempts",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"orgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
"shortName": "Baxter",
"dateUpdated": "2024-11-14T21:47:11.069Z"
},
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01"
}
],
"source": {
"discovery": "UNKNOWN"
},
"title": "No limit on failed login attempts with Clinician Password or Serial Number Clinician Password on Life2000 Ventilator",
"x_generator": {
"engine": "Vulnogram 0.2.0"
}
},
"adp": [
{
"affected": [
{
"vendor": "baxter",
"product": "life2000_ventilator_firmware",
"cpes": [
"cpe:2.3:o:baxter:life2000_ventilator_firmware:*:*:*:*:*:*:*:*"
],
"defaultStatus": "unaffected",
"versions": [
{
"version": "0",
"status": "affected",
"lessThanOrEqual": "06.08.00.00",
"versionType": "custom"
}
]
}
],
"metrics": [
{
"other": {
"type": "ssvc",
"content": {
"timestamp": "2024-11-18T15:36:22.221928Z",
"id": "CVE-2024-9832",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "total"
}
],
"role": "CISA Coordinator",
"version": "2.0.3"
}
}
}
],
"title": "CISA ADP Vulnrichment",
"providerMetadata": {
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP",
"dateUpdated": "2024-11-18T15:37:00.311Z"
}
}
]
}
}