Files
vuln-cve/data/2022/41xxx/CVE-2022-41627.json
T

166 lines
6.9 KiB
JSON
Raw Blame History

This file contains ambiguous Unicode characters
This file contains Unicode characters that might be confused with other characters. If you think that this is intentional, you can safely ignore this warning. Use the Escape button to reveal them.
{
"dataType": "CVE_RECORD",
"dataVersion": "5.1",
"cveMetadata": {
"cveId": "CVE-2022-41627",
"assignerOrgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"state": "PUBLISHED",
"assignerShortName": "icscert",
"requesterUserId": "bc31a57b-b1a5-40e2-9263-67c0ae8a3b8a",
"dateReserved": "2022-09-29T14:09:27.495Z",
"datePublished": "2022-10-27T20:04:06.498Z",
"dateUpdated": "2025-04-16T16:08:30.914Z"
},
"containers": {
"cna": {
"affected": [
{
"defaultStatus": "unaffected",
"product": "KardiaMobile",
"vendor": "AliveCor",
"versions": [
{
"status": "affected",
"version": "All"
}
]
}
],
"credits": [
{
"lang": "en",
"type": "finder",
"user": "00000000-0000-4000-9000-000000000000",
"value": "Carlos Cilleruelo Rodríguez"
},
{
"lang": "en",
"type": "finder",
"user": "00000000-0000-4000-9000-000000000000",
"value": "Javier Junquera Sánchez"
}
],
"datePublic": "2022-10-25T18:00:00.000Z",
"descriptions": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "\n\n<span style=\"background-color: rgb(255, 255, 255);\">The physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.</span>\n\n"
}
],
"value": "\nThe physical IoT device of the AliveCor's KardiaMobile, a smartphone-based personal electrocardiogram (EKG) has no encryption for its data-over-sound protocols. Exploiting this vulnerability could allow an attacker to read patient EKG results or create a denial-of-service condition by emitting sounds at similar frequencies as the device, disrupting the smartphone microphone’s ability to accurately read the data. To carry out this attack, the attacker must be close (less than 5 feet) to pick up and emit sound waves.\n\n"
}
],
"metrics": [
{
"cvssV3_1": {
"attackComplexity": "HIGH",
"attackVector": "PHYSICAL",
"availabilityImpact": "HIGH",
"baseScore": 4.8,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "NONE",
"integrityImpact": "LOW",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:N/S:U/C:N/I:L/A:H",
"version": "3.1"
},
"format": "CVSS",
"scenarios": [
{
"lang": "en",
"value": "GENERAL"
}
]
}
],
"problemTypes": [
{
"descriptions": [
{
"cweId": "CWE-311",
"description": "CWE-311 Missing Encryption of Sensitive Data",
"lang": "en",
"type": "CWE"
}
]
}
],
"providerMetadata": {
"orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6",
"shortName": "icscert",
"dateUpdated": "2022-11-01T15:57:59.910Z"
},
"references": [
{
"tags": [
"government-resource"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-298-01"
}
],
"source": {
"discovery": "EXTERNAL"
},
"x_generator": {
"engine": "cveClient/1.0.13"
}
},
"adp": [
{
"providerMetadata": {
"orgId": "af854a3a-2127-422b-91ae-364da2661108",
"shortName": "CVE",
"dateUpdated": "2024-08-03T12:49:43.451Z"
},
"title": "CVE Program Container",
"references": [
{
"tags": [
"government-resource",
"x_transferred"
],
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-298-01"
}
]
},
{
"metrics": [
{
"other": {
"type": "ssvc",
"content": {
"timestamp": "2025-04-16T15:49:32.342453Z",
"id": "CVE-2022-41627",
"options": [
{
"Exploitation": "none"
},
{
"Automatable": "no"
},
{
"Technical Impact": "partial"
}
],
"role": "CISA Coordinator",
"version": "2.0.3"
}
}
}
],
"title": "CISA ADP Vulnrichment",
"providerMetadata": {
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
"shortName": "CISA-ADP",
"dateUpdated": "2025-04-16T16:08:30.914Z"
}
}
]
}
}