220 lines
11 KiB
JSON
220 lines
11 KiB
JSON
{
|
||
"dataType": "CVE_RECORD",
|
||
"dataVersion": "5.1",
|
||
"cveMetadata": {
|
||
"cveId": "CVE-2022-26389",
|
||
"assignerOrgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
|
||
"state": "PUBLISHED",
|
||
"assignerShortName": "Baxter",
|
||
"dateReserved": "2022-03-03T22:04:24.534Z",
|
||
"datePublished": "2025-02-07T17:07:27.929Z",
|
||
"dateUpdated": "2025-02-07T18:49:22.506Z"
|
||
},
|
||
"containers": {
|
||
"cna": {
|
||
"affected": [
|
||
{
|
||
"defaultStatus": "unaffected",
|
||
"product": "ELI 380 Resting Electrocardiograph",
|
||
"vendor": "Baxter/ Hillrom",
|
||
"versions": [
|
||
{
|
||
"lessThanOrEqual": "2.6.0",
|
||
"status": "affected",
|
||
"version": "0",
|
||
"versionType": "custom"
|
||
}
|
||
]
|
||
},
|
||
{
|
||
"defaultStatus": "unaffected",
|
||
"product": "ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph",
|
||
"vendor": "Welch Allyn",
|
||
"versions": [
|
||
{
|
||
"lessThanOrEqual": "2.3.1",
|
||
"status": "affected",
|
||
"version": "0",
|
||
"versionType": "custom"
|
||
}
|
||
]
|
||
},
|
||
{
|
||
"defaultStatus": "unaffected",
|
||
"product": "ELI 250c/BUR 250c Resting Electrocardiograph",
|
||
"vendor": "Welch Allyn",
|
||
"versions": [
|
||
{
|
||
"lessThanOrEqual": "2.1.2",
|
||
"status": "affected",
|
||
"version": "0",
|
||
"versionType": "custom"
|
||
}
|
||
]
|
||
},
|
||
{
|
||
"defaultStatus": "unaffected",
|
||
"product": "ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph",
|
||
"vendor": "Welch Allyn",
|
||
"versions": [
|
||
{
|
||
"lessThanOrEqual": "2.2.0",
|
||
"status": "affected",
|
||
"version": "0",
|
||
"versionType": "custom"
|
||
}
|
||
]
|
||
}
|
||
],
|
||
"credits": [
|
||
{
|
||
"lang": "en",
|
||
"type": "finder",
|
||
"value": "An anonymous user reported these vulnerabilities to Hillrom."
|
||
}
|
||
],
|
||
"descriptions": [
|
||
{
|
||
"lang": "en",
|
||
"supportingMedia": [
|
||
{
|
||
"base64": false,
|
||
"type": "text/html",
|
||
"value": "An improper access control vulnerability may allow privilege escalation.<p>This issue affects: </p><ul><li>ELI 380 Resting Electrocardiograph:\n\n<span style=\"background-color: rgb(255, 255, 255);\">Versions 2.6.0 and prior;</span> </li><li>ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph:\n\n<span style=\"background-color: rgb(255, 255, 255);\">Versions 2.3.1 and prior</span>; </li><li>ELI 250c/BUR 250c Resting Electrocardiograph: <span style=\"background-color: rgb(255, 255, 255);\">Versions 2.1.2 and prior; </span></li><li>ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: \n\n<span style=\"background-color: rgb(255, 255, 255);\">Versions 2.2.0 and prior</span>.</li></ul>"
|
||
}
|
||
],
|
||
"value": "An improper access control vulnerability may allow privilege escalation.This issue affects: \n\n * ELI 380 Resting Electrocardiograph:\n\nVersions 2.6.0 and prior; \n * ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph:\n\nVersions 2.3.1 and prior; \n * ELI 250c/BUR 250c Resting Electrocardiograph: Versions 2.1.2 and prior; \n * ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: \n\nVersions 2.2.0 and prior."
|
||
}
|
||
],
|
||
"impacts": [
|
||
{
|
||
"capecId": "CAPEC-233",
|
||
"descriptions": [
|
||
{
|
||
"lang": "en",
|
||
"value": "CAPEC-233 Privilege Escalation"
|
||
}
|
||
]
|
||
}
|
||
],
|
||
"metrics": [
|
||
{
|
||
"cvssV3_1": {
|
||
"attackComplexity": "HIGH",
|
||
"attackVector": "NETWORK",
|
||
"availabilityImpact": "HIGH",
|
||
"baseScore": 7.7,
|
||
"baseSeverity": "HIGH",
|
||
"confidentialityImpact": "LOW",
|
||
"integrityImpact": "LOW",
|
||
"privilegesRequired": "LOW",
|
||
"scope": "CHANGED",
|
||
"userInteraction": "NONE",
|
||
"vectorString": "CVSS:3.1/AV:N/AC:H/PR:L/UI:N/S:C/C:L/I:L/A:H",
|
||
"version": "3.1"
|
||
},
|
||
"format": "CVSS",
|
||
"scenarios": [
|
||
{
|
||
"lang": "en",
|
||
"value": "GENERAL"
|
||
}
|
||
]
|
||
}
|
||
],
|
||
"problemTypes": [
|
||
{
|
||
"descriptions": [
|
||
{
|
||
"cweId": "CWE-284",
|
||
"description": "CWE-284 Improper Access Control",
|
||
"lang": "en",
|
||
"type": "CWE"
|
||
}
|
||
]
|
||
}
|
||
],
|
||
"providerMetadata": {
|
||
"orgId": "dba971b9-eb30-4121-91e1-3b45611354aa",
|
||
"shortName": "Baxter",
|
||
"dateUpdated": "2025-02-07T17:07:27.929Z"
|
||
},
|
||
"references": [
|
||
{
|
||
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-22-167-01"
|
||
},
|
||
{
|
||
"url": "https://hillrom.com/en/responsible-disclosures/"
|
||
}
|
||
],
|
||
"solutions": [
|
||
{
|
||
"lang": "en",
|
||
"supportingMedia": [
|
||
{
|
||
"base64": false,
|
||
"type": "text/html",
|
||
"value": "<p>Hillrom has released software updates for all impacted devices to \naddress these vulnerabilities. New product versions that mitigate these \nvulnerabilities are available as follows:</p>\n<ul>\n<li>Welch Allyn ELI 380 Resting Electrocardiograph: available Q4 2023</li>\n<li>Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: available May 2022</li>\n<li>Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: available Q4 2023</li>\n</ul>\n<p>Hillrom recommends users upgrade to the latest product versions. \nInformation on how to update these products can be found on the <a target=\"_blank\" rel=\"nofollow\" href=\"https://hillrom.com/en/responsible-disclosures/\">Hillrom disclosure page</a></p>.\n\n<br>"
|
||
}
|
||
],
|
||
"value": "Hillrom has released software updates for all impacted devices to \naddress these vulnerabilities. New product versions that mitigate these \nvulnerabilities are available as follows:\n\n\n\n * Welch Allyn ELI 380 Resting Electrocardiograph: available Q4 2023\n\n * Welch Allyn ELI 280/BUR280/MLBUR 280 Resting Electrocardiograph: available May 2022\n\n * Welch Allyn ELI 150c/BUR 150c/MLBUR 150c Resting Electrocardiograph: available Q4 2023\n\n\n\n\nHillrom recommends users upgrade to the latest product versions. \nInformation on how to update these products can be found on the Hillrom disclosure page https://hillrom.com/en/responsible-disclosures/ \n\n."
|
||
}
|
||
],
|
||
"source": {
|
||
"advisory": "ICSMA-22-167-01",
|
||
"discovery": "EXTERNAL"
|
||
},
|
||
"title": "Improper Access Control Vulnerability in ELI Electrocardiograph Devices",
|
||
"workarounds": [
|
||
{
|
||
"lang": "en",
|
||
"supportingMedia": [
|
||
{
|
||
"base64": false,
|
||
"type": "text/html",
|
||
"value": "<p>Hillrom recommends the following workarounds to help reduce risk:</p>\n<ul>\n<li>Apply proper network and physical security controls.</li>\n<li>Ensure a unique encryption key is configured for ELI Link and Cardiograph.</li>\n<li>Where possible, use a firewall to prevent communication on Port 21 \nFTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet \nservice.</li>\n</ul>\n\n<br>"
|
||
}
|
||
],
|
||
"value": "Hillrom recommends the following workarounds to help reduce risk:\n\n\n\n * Apply proper network and physical security controls.\n\n * Ensure a unique encryption key is configured for ELI Link and Cardiograph.\n\n * Where possible, use a firewall to prevent communication on Port 21 \nFTP service, Port 22 SSH (Secure Shell Connection), and Port 23 Telnet \nservice."
|
||
}
|
||
],
|
||
"x_generator": {
|
||
"engine": "Vulnogram 0.2.0"
|
||
}
|
||
},
|
||
"adp": [
|
||
{
|
||
"metrics": [
|
||
{
|
||
"other": {
|
||
"type": "ssvc",
|
||
"content": {
|
||
"timestamp": "2025-02-07T18:49:13.343999Z",
|
||
"id": "CVE-2022-26389",
|
||
"options": [
|
||
{
|
||
"Exploitation": "none"
|
||
},
|
||
{
|
||
"Automatable": "no"
|
||
},
|
||
{
|
||
"Technical Impact": "partial"
|
||
}
|
||
],
|
||
"role": "CISA Coordinator",
|
||
"version": "2.0.3"
|
||
}
|
||
}
|
||
}
|
||
],
|
||
"title": "CISA ADP Vulnrichment",
|
||
"providerMetadata": {
|
||
"orgId": "134c704f-9b21-4f2e-91b3-4a467353bcc0",
|
||
"shortName": "CISA-ADP",
|
||
"dateUpdated": "2025-02-07T18:49:22.506Z"
|
||
}
|
||
}
|
||
]
|
||
}
|
||
} |