{ "containers": { "cna": { "affected": [ { "defaultStatus": "unaffected", "product": "Conexus Radio Frequency Telemetry Protocol", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "MyCareLink Monitor", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "24950" }, { "status": "affected", "version": "24952" } ] }, { "defaultStatus": "unaffected", "product": "CareLink Monitor", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "2490C" } ] }, { "defaultStatus": "unaffected", "product": "CareLink 2090 Programmer", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Amplia CRT-D", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Claria CRT-D", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Compia CRT-D", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Concerto CRT-D", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Concerto II CRT-D", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Consulta CRT-D", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Evera ICD", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Maximo II CRT-D", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Maximo II ICD", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Mirro ICD", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Nayamed ND ICD", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Primo ICD", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Protecta ICD, Protecta CRT-D", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Secura ICD", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Virtuoso ICD", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Virtuoso II ICD", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Visia AF ICD", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Viva CRT-D", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Brava CRT-D", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] }, { "defaultStatus": "unaffected", "product": "Mirro MRI ICD", "vendor": "Medtronic", "versions": [ { "status": "affected", "version": "All versions" } ] } ], "credits": [ { "lang": "en", "type": "finder", "value": "Peter Morgan of Clever Security; Dave Singelée and Bart Preneel of KU Leuven; Eduard Marin formerly of KU Leuven, currently with University of Birmingham; Flavio D. Garcia; Tom Chothia of the University of Birmingham; and Rik Willems of University Hospital Gasthuisberg Leuven reported these vulnerabilities to CISA." } ], "datePublic": "2019-03-21T06:00:00.000Z", "descriptions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "
The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data.
" } ], "value": "The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement encryption. An attacker with adjacent short-range access to a target product can listen to communications, including the transmission of sensitive data." } ], "metrics": [ { "cvssV3_1": { "attackComplexity": "LOW", "attackVector": "ADJACENT_NETWORK", "availabilityImpact": "NONE", "baseScore": 6.5, "baseSeverity": "MEDIUM", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N", "version": "3.1" }, "format": "CVSS", "scenarios": [ { "lang": "en", "value": "GENERAL" } ] } ], "problemTypes": [ { "descriptions": [ { "cweId": "CWE-319", "description": "CWE-319 Cleartext Transmission of Sensitive Information", "lang": "en", "type": "CWE" } ] } ], "providerMetadata": { "orgId": "7d14cffa-0d7d-4270-9dc0-52cabd5a23a6", "shortName": "icscert", "dateUpdated": "2025-05-22T19:23:42.770Z" }, "references": [ { "tags": [ "x_refsource_MISC" ], "url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01" }, { "name": "107544", "tags": [ "vdb-entry", "x_refsource_BID" ], "url": "http://www.securityfocus.com/bid/107544" } ], "solutions": [ { "lang": "en", "supportingMedia": [ { "base64": false, "type": "text/html", "value": "Medtronic has developed mitigating patches for a subset of the \naffected implanted cardiac device models. These patches are installed \nduring regular office visits. Medtronic has stated that patches for \nadditional impacted models are being developed by Medtronic and will be \ndeployed through future updates. Patches are currently available for the\n following affected models:
Medtronic has applied additional controls for monitoring and responding to improper use of the Conexus telemetry protocol by the affected implanted cardiac devices. Additional mitigations are being developed and will be deployed through future updates, assuming regulatory approval.
Medtronic recommends that users take additional defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:
Medtronic has released additional patient focused information.